42
Operation
Treatment STANDBY mode
After the system has been switched on and the internal system test has
been completed, the system enters this mode automatically unless the
instrument is switched to Diagnosis mode by the user.
Fig. 22
Treatment STANDBY mode
All treatment-related laser parameters are displayed on the control
panel and can be changed in this mode.
The following laser parameters can be set in this mode: energy, pulse
mode and aiming beam intensity.
Each time the system is restarted, the energy and the aiming beam
intensity values most recently selected will be used.
• To set individual laser parameters, touch the appropriate parameter
button and set the value required by turning the control knob.
• Select the desired pulse mode (single, double or triple pulse) by
touching the appropriate button. The reverse-coloured button (blue
on a white background) is activated.
• Pressing the control knob takes you directly to the entry field for the
laser energy. Rotate the knob to set the energy required.
000000-1354-664 VISULAS YAG III 15.08.2005