Operation

READY

Fig. 33
Treatment READY mode
In Treatment READY mode, the aiming beam and laser warning
lamp are always on.
Parameters are set in the same way as for Treatment STANDBY
mode.
When the firing switch is activated, the laser beam is fired according
to the parameters set.
Touching the
STANDBY/READY
into Treatment STANDBY mode.
Touching the
ARROW
Note
If, in Treatment READY mode, more than five minutes
elapses without the laser being fired, the system returns to
Treatment STANDBY mode and emits a short beep.
button switches the system back
button returns you to the Main menu.