58
An error has occurred on the
VISULAS YAG
Error messages and information on the VISULAS 532s control unit
Message
Door interlock is active, please close
door.
LASER STOP is activated; to continue,
please unlock.
No physician safety filter or slit lamp
connected!
No fiber connected!
No foot switch connected!
Temperature deviation, please wait!
Please swivel lamp housing.
Pulse interval matched.
More than 20 % power deviation.
The foot switch is activated.
Maintenance / Miscellaneous
Faults during operation of the VISULAS YAG III Combi
If, in YAG mode, an error occurs on the VISULAS YAG III, an error
message will be displayed on the control panel.
At the same time, an error message with a specific error code will be
displayed on the YAG console.
Remedy the fault according to the table on page 57, then switch the
VISULAS 532s off and then on again.
Faults on the VISULAS 532s are indicated as system information on the
control unit display. The user is also alerted by short beeps.
Note:
Warning message:
If a fault occurs, which cannot be remedied by using the following
tables, please contact a Carl Zeiss Meditec authorised specialist, citing
the error number displayed.
Label the instrument as being out of order and remove the key from
the key switch.
Remedy
Shut the door or check that the shorting plug is correctly plugged into the yellow
socket.
Unlock the LASER STOP button.
Check that the slit lamp and the OPMI (ENDO physician's safety filter) are
correctly connected and properly inserted.
Check that the optical cable is properly inserted and screwed in tightly.
Check that the foot switch is correctly connected and properly inserted.
If this fault occurs frequently, ensure that the instrument's ventilation slots are
not covered and that the ambient temperature falls below 35 °C.
Swivel the lamp casing out of position. Check that all the connectors on the rear
of both consoles are inserted properly.
If the laser parameters 'pulse length or 'power' are changed, the system may
carry out an adjustment of the 'pulse interval' parameter. To avoid overloading
the system, the interval between laser pulses is increased.
The internal monitoring systems have identified that the actual laser power
output deviates from the target power output selected on the control panel by
more than 20 %. Should this error message occur repeatedly, please inform our
service department.
Release the foot switch.
Follow the instructions given.
Remedy
the
fault
troubleshooting tables below.