Maintenance / Miscellaneous
• Adjust the ENERGY value displayed on the control unit until it lies
within the tolerance range. To do so, use the potentiometer on the
square PCB at the top of the laser head (towards the binocular
tube):
Turn the potentiometer:
clockwise:
counter clockwise:
• Fire five laser shots and calculate the new mean. Readjust the
potentiometer if necessary.
• Repeat this process until the displayed ENERGY is within the
tolerance range.
• Reattach the cover and remount the focus shift control.
to increase the ENERGY
to reduce the ENERGY