2.2.1.

HF device

Use only approved HF devices and programs (see
section section 8, page 31).
Observe the operating instructions of the HF device and the
general instructions for electrosurgical operations!
Improper use of HF current can lead to endogenous and exogenous
burns and explosions:
Avoid direct skin contact with HF cables.
Avoid contact with flammable gases and liquids.
2.2.2.

HF cable

Improper use of HF cables can lead to patient injuries:
Never lay the HF cable on the patient's skin.
Connect the ligation instrument for coagulation before switching
on the HF generator.
When plugging or unplugging the HF cable, always grasp the
connector directly.
Use only HF cables that are in perfect condition. Never use a
defective HF cable.
The HF cable may cause interference to imagery on monitors:
Never route the HF cable alongside a camera cable.
Do not lay the HF cable in loops.
Consult the operating manuals of the BOWA HF generators for
additional information on interference with other devices.
MN031-543-S0 EN
TissueSeal / TissueSeal PLUS COMFORT
2 Safety
13