7 Preparation
7.2.

Dismantling

Dismantle the ligation instrument (see section 6, page 19).
7.2.1.

Pretreatment in an ultrasonic bath

Risk of infection due to water spray and vapors from the
ultrasonic bath!
1. Place the ligation instrument in the ultrasonic bath for at least 5 minutes.
Position instrument parts with large surface areas in the ultrasonic bath
such that they will not be damaged by the ultrasonic energy.
2. Use suitable cleaning and disinfection products for ultrasonic cleaning
(see section 7.11, page 30).
3. Follow the manufacturer's instructions with regard to the concentration
and exposure time of the cleaning and disinfection products.
24
TissueSeal / TissueSeal PLUS COMFORT
CAUTION
Wear a face mask and protective clothing.
Adequate ventilation is recommended.
MN031-543-S0 EN