Transport; Functional Test In The Operating Room; Recommended Operating Supplies - Bowa TissueSeal PLUS COMFORT Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 149
7 Preparation
7.9.

Transport

Strictly observe the storage conditions when transporting the
ligation instrument.
Transport conditions:
Temperature:
Relative humidity:
Atmospheric pressure:
7.10th

Functional test in the operating room

1.
Assemble the ligation instrument (see section 4.1, page 15).
2.
Test the ligation instrument for proper operation (see section 7.5,
page 26.
7.11.

Recommended operating supplies

BOWA recommends the use of neutral to slightly alkaline cleaning
agents or cleaning and disinfection agents free from potentially
harmful ingredients. Alcoholic and/or aldehydic ingredients may be
permissible, depending on the concentration.
Pretreatment in an ultrasonic bath
The suitability of the ligation instruments for effective pretreatment in
an ultrasonic bath (5 minutes) with the use of an aldehyde-free
combined cleaning and disinfection product (Gigasept Instru AF) has
been demonstrated by BOWA.
Automatic cleaning
The suitability of the ligation instruments for effective cleaning and/or
disinfection with an automated method (90 °C, 5 minutes) and an
alkali cleaning product with a surfactant additive (neodic MediCLean
forte) has been demonstrated by BOWA.
The manufacturer accepts no responsibility for the use of other
cleaning and disinfection products.
32
TissueSeal / TissueSeal PLUS COMFORT
- 20 °C - + 50 °C
0 – 75%
500 – 1600 hPa
MN031-543-S0 EN

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tissueseal comfort

Tabla de contenido