Inspection
Packing
Autoclaving
Storage
Functional test in the operating room
7.1.
Soaking
Risk of infection due to water spray and vapors from the
ultrasonic bath or with manual cleaning!
Risk of material damage to the ligation instrument from
scouring agents and metal brushes!
If necessary, first use a nonwoven cloth to remove residual dirt.
Soak the ligation instrument immediately after use, or no later
than 2 hours after use.
Use only aldehyde-free disinfectants suitable for disinfecting
ligation instruments, such DGHM- or FDA-approved products or
products with the CE mark.
The disinfectant used for soaking is intended solely for
personal protection and does not replace subsequent
disinfection.
MN031-543-S0 EN
CAUTION
Wear a face mask and protective clothing.
Adequate ventilation is recommended.
NOTE
Never use scouring products to clean the ligation
instrument.
TissueSeal / TissueSeal PLUS COMFORT
7 Preparation
23