Risk of patient injury due to hot electrode surfaces and
vapor emission!
Grasping, clamping and sealing tissue
Risk of patient injury due to inadvertent activation of the
ligation instrument!
1.
Place the jaws A on the operating site.
2.
Place the tissue to be sealed between the electrodes of the jaws
A.
3.
Close the jaws A to grasp the tissue. Do not grasp too much
tissue.
The tissue is grasped.
4.
Use the two-position ratchet of the handle B to adjust the
pressure on the tissue to best match the amount of grasped
tissue.
The tissue is clamped.
MN031-543-S0 EN
WARNING
Maintain sufficient distance between the tips of the
instrument and sensitive tissue structures, such as the
pancreas or intestine.
Ensure that hot ligation instruments are not used for
preparation.
Do not lay the ligation instrument on the patient.
WARNING
Never use the AUTOSTART function.
Do not switch on the HF current before the active
electrodes are in contact with the tissue to be
coagulated and the ratchet is closed.
TissueSeal / TissueSeal PLUS COMFORT
5 Operation
19