Water leakage between the handpiece
cord and the handpiece
Water leakage between the handpiece
and the nozzle
Air and powder leakage on the main unit
Air and powder is leaking at powder
chamber
decreases during use
Leakage on the unit water supply
•
Check the condition of the handpiece connecting O-rings
•
Send the system and the handpiece cord to an authorized EMS Repair Center
•
Verify the position of the nozzle and its position on the handpiece
•
Verify that no residual remains between the nozzle and the handpiece
•
•
Replace the nozzle
•
Replace the handpiece
•
•
Verify the positioning of the bottles
•
Check the O-rings under the silicone caps of the powder chamber holders
•
Verify the cap of the powder chamber and the O-ring positioning
•
Verify the check valve and the O-ring under the powder chamber
•
Remove all powder dust between the bottle holder and the powder chamber
•
Send the unit to an authorized EMS Repair Center
•
Check the seal condition and the cleanness of the thread of the powder chamber cap
•
Replace the seals if necessary
•
Clean the sealing surfaces carefully
•
Replace the powder chambers
•
Send the unit to an authorized EMS Repair Center
•
Verify the quantity of powder in your unit
•
Eliminate residual powder present in the channels using a cleaning needle
•
•
Check the water supply connection and inlet pressure
•
Send the unit to an authorized EMS Repair Center
93