5. Remove from washer and dry with a clean, soft towel.
6. Inspect all components to ensure there is no visible organic debris or residue from cleaning
agent. Repeat process if any visible soil is detected.
(Pre-Wash)
Temperature Hold Time
Cold water from
building supply
93°C
(less than 110° Or
43°C)/4 min.
50°C/7 min
90°C
Optional Thermal Rinse:
• A Thermal Disinfection rinse may be added after the first rinse cycle as indicated in Table 3.
Additionally Validated Cleaning Cycles
Wash
Detergent
Treatment
Alkaline
1 or more
(e.g. neodisher®
min.
SeptoClean 1%)
Neutral Enzymatic
(e.g. Enzol® per
1 or more
manufacturer's
min.
recommendation)
Table 4 Additionally Validated Cleaning Cycles
19
Thermal
(Rinsing)
Disinfection
Hot water
from building
Optional
supply/2 min.
Hot water
from building
Optional
supply/2 min.
EN – ENGLISH
(Drying)
(Optional)
90°C/2O min.
90°C/20 min.