Manual Cleaning
1. Prepare a fresh solution of detergent with lukewarm water.
Note: see the Materials and Equipment table for validated
detergents with their minimum concentration and soak time.
2. Wipe the entire surface of the device using a soft clean cloth
dipped in the detergent solution.
3. Immerse the device in the detergent solution, ensuring the
solution contacts all inner and outer surfaces.
4. Soak the device in the solution according to the manufacturer's
recommendations.
5. With the device immersed in the solution, thoroughly brush the
exterior with a soft-bristle brush, focusing on any mated or rough
surfaces.
6. Actuate and brush any movable parts in all extreme positions.
7. Rinse the device with water until all detergent residue is removed.
8. Once all detergent residue is removed, continue to rinse for 30
seconds.
9. Drain excess water from the device and dry it using a clean cloth or
pressurized air.
10. Visually inspect the device for cleanliness, paying close attention to
hard-to-reach areas. If visible soil remains, repeat steps 1– 10.
Automated Cleaning
The device cannot withstand an automated disinfection
method. When programming the washer, do not include a
thermal rinse cycle or the device will be damaged.
1. Place the device in the automated washer on an incline to facilitate
drainage.
2. Program the washer using the following parameters:
Phase
Pre-Wash
Wash 1
Rinse 1
Dry Phase
3. Drain excess water from the device and dry it using a clean cloth or
pressurized air.
4. Visually inspect each device for cleanliness, paying close attention
to hard-to-reach areas. If visible soil remains, repeat steps 1– 4.
Recirculation
Time
2 minutes
5 minutes
2 minutes
2 minutes
Temperature
Cold Water
Set Point 60 °C
(140 °F)
Hot Water
115 °C (239 °F)
Detergent
Type
N/A
See Materials
and
Equipment
table
N/A
N/A