Product / User / Disposal:
Electrosurgery accessories may only be used and disposed of by qualified medical personnel! Dispose of the
instrument according to internal hospital guidelines for sharp, biologically contaminated items.
Notwithstanding these instructions, reading the instructions for use provided with the electrosurgery device
and other accessories being used is mandatory.
a   N ot sterile. Must be cleaned and sterilized prior to initial and each subsequent use.
Intended purpose:
Sutter SuperGliss
, SuperGliss
ELP, SuperGliss
®
®
tal™ bipolar forceps for coagulation of the selected tissue.
Sutter bipolar suction forceps for use in electrosurgery for the coagulation of tissue and suctioning fluids.
Sutter bipolar irrigation forceps for use in electrosurgery for coagulation and to supply fluid to select tissue.
Service life:
A minimum of 20 reconditioning cycles can be expected with proper use.
TEO, SuperGliss
zhora, Classic, Classic Micro and Selec-
®
®
Before using:
EN
a   I nspect the product for cleanliness, mechanical function, and intact insulation before each use.
We recommend checking the insulation with a suitable test device.
a   O nly use sterilized products that are in flawless condition!
A certain discoloration of the non-stick instrument tip is normal and harmless.
To connect the forceps and cables, the electrosurgery device must be turned off or in standby mode. Failure
to comply may lead to burns and electric shocks!
For further information about electrical safety, we recommend DIN EN 60601-2-2 supplement 1.
Electrosurgery cables:
The Sutter bipolar forceps are intended for use with bipolar silicone cables with US male connector or
European flat connector from the manufacturer Sutter Medizintechnik GmbH.
During use:
a   A lways work with the lowest output setting for the desired surgical effect.
a   M aximum allowable voltage 500 Vp.
a   R egularly wipe blood and tissue residue from the tips.
a   T ips of the forceps may cause injuries!