Ottobock 6A20-10 Instrucciones De Uso página 10

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27
prosthetic socket. Then the shuttle lock is screwed into the lamination
anchor.
1.2 Combination possibilities
This prosthetic component is compatible with Ottobock's system of
modular connectors. Functionality with components of other manufac­
turers that have compatible modular connectors has not been tested.
2 Intended use
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
The maximum approved body weight is specified in the technical
data (see Page 15).
2.3 Environmental conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service life
Depending on the patient's level of activity, the service life of the
product is 3 to 5 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or
CAUTION
injury.
10
3.2 General safety instructions
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
► Use the product according to the specified area of application
(see Page 10).
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
► Only combine the product with prosthetic components that are
approved for that purpose.
► Based on the instructions for use of the prosthetic components,
verify that they may be combined with each other.
CAUTION
Use under unallowable environmental conditions
Risk of injury due to damage to the product
► Do not expose the product to unallowable environmental condi­
tions.
► If the product has been exposed to unallowable environmental
conditions, check it for damage.
► If damage is apparent or in case of doubt, do not continue using
the product.
► Take suitable measures if required (e.g. cleaning, repair,
replacement, inspection by the manufacturer or a specialist
workshop etc.).
CAUTION
Exceeding the service life and reuse on another patient
Risk of injury due to loss of functionality as well as damage to the
product

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

6a20-20

Tabla de contenido