9. Warranty and Maintenance
IlluminOss Medical Inc offers a one year warranty against defects in material and workmanship on all
system components, except the bulb. Unauthorized repair, modification, or improper use of equipment will
void the warranty. The use of aftermarket replacement parts not supplied or approved by IlluminOss
Medical Inc will void any effective warranties and may result in damage to the equipment.
The IlluminOss Medical Inc Light Source console was designed to operate with minimal maintenance.
Follow the listing below to assure top unit performance. In addition, examine cords, connections, and
console for wear or damage monthly. After every use, clean the Light Source Console, Light Guide, Foot
Pedal, and power cord according to the Cleaning and Sterilization section of this manual.
Clean the Light Guide ends monthly or as required according to the Cleaning and Sterilization section of this
manual. The ends of the Guide should be kept clean to transmit as much light as possible. Handle the Light
Guide with care, as sharp bends will reduce light output and cause permanent damage.
The external fan filters should be inspected and cleaned
periodically to prevent dust build up from affecting
airflow through the unit. Spare filters are provided with
each unit and with replacement bulbs. The fan filters are
washable and may be reused. Replace the fan filter by
removing the snap-on cover from the rear of each grill.
F
R
USE
EPLACEMENT
Contact IlluminOss for servicing.
Clean the Photodynamic Curing system immediately after each use by following the cleaning procedure
described. These procedures require the use of a low-level disinfectant (LLD) wipe and solution that
includes the following claims on its labeling – that it does not cause adverse effects (corrosion) to plastics,
rubbers, and metals.
Users of the Photodynamic Light Curing System must also observe site-specific governing regulations for
protection of personnel and for effective handling and disposal of LLD waste by-products.
CAUTION: DO NOT IMMERSE THE LIGHT SOURCE CONSOLE, LIGHT GUIDE, POWER CORD, OR FOOT PEDAL
IN ANY SOLUTION, OR CLEAN WITH STEAM, AUTOCLAVE OR ETHYLENE OXIDE.
900093.W issued 01-JUL-21
W
ARRANTY
M
AINTENANCE
L
G
IGHT
UIDE
F
AN
B
ULB REPLACEMENT
The bulb requires replacement after 500 hours of use. After 450
hours of use, the red "Change Bulb" indicator light on the front
panel of the box will flash to warn the user of the impending need
to replace the bulb. The unit will be operational while the light is
flashing. After 500 hours, the red "Change Bulb" light will
permanently illuminate, there will be an audible beeping sound
and the unit will be inoperable. Return the unit to IlluminOss for
bulb replacement when the indicator light starts flashing.