WARNING:
WARNING:
The power supply operates on line voltage between 100-240 VAC and frequencies between 50-60 Hz. If the
bulb extinguishes due to a momentary power failure, the unit must be turned off, allowed to cool, and then
turned on again in order to reignite the bulb. A cooling fan is provided to keep the bulb housing and internal
components of the power supply at the optimum operating temperature. The fan must not be covered or
otherwise blocked. The light source is a 75-Watt short arc mercury vapor bulb mounted in a reflector and
pre-focused to provide optimum light output. The unit is rated for continuous operation.
The system is designed to operate in conjunction with the IlluminOss Variable Timer Keys which will be
supplied with each IlluminOss Photodynamic Bone Stabilization System. The keys contain specific resistor
circuitry which, when connected to the light box, will program a specific timer countdown cycle which is
optimized for the specific IlluminOss implant.
Testing confirms a minimum acceptable intensity provided by the lamp bulb. A "Change Bulb" Indicator light
is provided to alert the operator to check bulb operation or to contact customer service if maintenance is
required.
900093.W issued 01-JUL-21
Engage the light guide in the mount before the light is turned on, and
remove the light guide from the bezel only after the light is turned off
to avoid the accidental possibility of exposure to the light.
UV emitted from this product.
F
P
RONT
PHOTODYNAMIC LIGHT CURING SYSTEM INSTRUCTIONS FOR USE
L
D
ANEL
IGHT
ISPLAY
SYSTEM READY
LIGHT GUIDE SEATED
CHANGE BULB
PAGE 7 OF 121
-