8. Operation

The Photodynamic Light Curing System will arrive almost fully assembled. The system should be positioned
in a dry location that will not obstruct the rear of the unit. Adequate ventilation is essential for proper
operation.
CAUTION: TO ENSURE PROPER OPERATION OF THE SHUTTER, BE SURE TO COMPLETELY INSERT THE LIGHT
GUIDE INTO THE CONSOLE. ENSURE THAT THE LIGHT GUIDE REMAINS IN PLACE DURING USE.
To power the system, turn the Power Switch to the ON position; the fan, timer, and 75W lamp should begin
to function. The operator should confirm proper start up before commencing with surgery by checking for
an audible fan system and visually confirming a green light shining out the back of the box. Before operating
the unit, allow the 75W lamp to complete the warm up cycle, approximately five minutes.
CAUTION: NEVER LOOK DIRECTLY AT LIGHT EXITING THE LIGHT GUIDE OR THE LIGHT PIPE.
CAUTION: IF POWER TO THE CONSOLE IS DISTRUPTED DURING A CURING CYCLE, TO ENSURE PROPER CURING
HAS BEEN ACHIEVED, IT IS RECOMMENDED THAT A FULL REPEAT CYCLE IS PERFORMED FOLLOWING THE
RESUMPTION OF POWER AND A FULL SYSTEM RESTART. THE CYCLE TIME IS AUTOMATICALLY SET BY THE
TIMER KEY INCLUDED IN THE CATHETER PACKAGE.
CAUTION: WHEN THE "CHANGE BULB" LIGHT ILLUMINATES, THE UNIT WILL NOT OPERATE UNTIL A NEW
BULB IS INSTALLED AS DESCRIBED IN THE BULB REPLACEMENT PROCEDURE.
Bulb life is reduced each time it is started. To avoid premature lamp deterioration, leave the unit on
through- out breaks and short shutdowns. These light sources are designed for continuous operation. If the
power is lost momentarily, shut the unit off to allow the unit to cool down for 5 -10 minutes, then restart
the console.
Turn the power switch to "ON." The LED display
should run through a power up sequence and end
on 2000 seconds.
Allow the bulb to warm up (3-5 minutes). The
bulb is warmed up when the GREEN "System-
Ready" indicating light is illuminated. The shutter
is inoperable until the warm up period is
complete and the Light Guide is installed.
900093.W issued 01-JUL-21
PHOTODYNAMIC LIGHT CURING SYSTEM INSTRUCTIONS FOR USE