2.2. OPERATING GUIDELINES
The myray RX DC - RX DC/I can function in different modes (MULTIMODE technology):
1) En60 operating mode
X-ray emission at 60KV and 7ma. The MyRay RX DC x-ray unit automatically calculates the best exposure time
(from 0.01s to 1.00s) based on the selected tooth and patient size.
2) En63 operating mode
X-ray emission at 63KV and 6ma. The MyRay RX DC x-ray unit automatically calculates the best exposure time
(from 0.01s to 1.00s) based on the selected tooth and patient size.
3) En65 operating mode
X-ray emission at 65KV and 6ma. The MyRay RX DC x-ray unit automatically calculates the best exposure time
(from 0.01s to 1.00s) based on the selected tooth and patient size.
4) AUTO operating mode
The MyRay RX DC x-ray unit automatically suggests the best operating mode (En60, En63 o En65) and
exposure time (from 0.01s to 1.00s) based on the selected tooth and patient size.
The suggested exposure time can be corrected from the handheld.
In addition, the MyRay RX DC x-ray unit has a special USER mode that can be set by the user. The user can select
the best combination of load factors (operating mode and exposure time) for each tooth and patient size.
3. FUNCTIONING
3.1. SWITCHING ON AND OFF THE X-RAY DEVICE
3.1.1. TURNING ON THE X-RAY UNIT
NOTE: The technical specifications of the switch are given in paragraph 1.2.2.
Whenever turned on, the equipment performs an operational test that takes a few seconds. A beep is
provided at the end of the test.
NOTE: The exposure time and the parameters displayed when the unit is turned on are the last ones set
before the central control unit was turned off.
If the central control unit is left untouched for a few minutes it will go into standby mode. Simply press any
key on the control panel to reactivate it.
EN
Wall mounted version (RX DC Plus and RX DC eXTend).
The control unit is turned on and shut off from the main
switch (A).
The switch lights up to signal the control unit is energized.
INCORPORATED version (RX DC PLUS / I).
The control unit is automatically turned on and off from the
main switch of the dental unit.
However, the x-ray unit has its own power supply line with
wall-mounted switch (not provided) used to shut off power
separately from the dental unit.
Make sure the switch has been turned on when the
apparatus needs to be used.
OPERATOR'S MANUAL