Error Messages; Periodic Maintenance - Myray RX DC Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 159

6. ERROR MESSAGES

NOTE: In presence of strong wireless communication traffic, the connection between handheld and
generator may break down. To restore the connection run the "Restoring factory settings" procedure.
ERROR
E01
X-RAY KEY RELEASED TOO EARLY
E02
SHOOTING SEQUENCE NOT COMPLETED
E03
HANDHELD INTERNAL TEST ERROR
E04
HANDHELD
E05
FAILED
E08
E06
GENERAL HANDHELD ERROR
E07
RF SIGNAL TOO LOW
HANDHELD SERIAL NUMBER INCORRECT
E09
OR NOT INITIALIZED
E10
E12
X-RAY UNIT INTERNAL ERROR
E13
E16
COLLIMATOR
E11
CONSISTENT
E14
GENERAL GENERATOR ERROR
E15
E17
DEVICE OVERHEATING
E18
HEAD RELEASED
E19
E30
SUPPLY VOLTAGE TOO HIGH/LOW
E31
INTERNAL ADJUSTMENT PROBLEM
E32
E33
REMOTE CONTROL ERROR
NOTE: As regards the other error codes, CONTACT the technical service department.

7. PERIODIC MAINTENANCE

WARNING:
Any technical maintenance work required must be carried out by qualified personnel or by a specialised
technician authorised by the manufacturer. It is the user's responsibility to check that routine maintenance is
carried out by an authorised technician at least every 2 years. The maintenance methods are specified in
the Technical Service Manual possessed by the Authorised Technicians.
74
CAUSE
AUTO
DIAGNOSIS
SELECTION
OPERATOR'S MANUAL
Hold down the key until the image has been
captured.
Handheld most likely lost the signal. Try to repeat
exposure.
If
technical service.
Take out the batteries and then put them back in
after waiting a few seconds. If the problem
persists, contact technical service.
TEST
Contact technical service department.
Try to repeat exposure. If the problem persists,
contact the technical service department.
Handheld lost the signal. Try to repeat exposure.
If the problem persists, contact the technical
service department.
Contact technical service department.
Contact technical service department
After turning the rectangular collimator on or off,
NOT
wait a few seconds to allow the icon on the
handheld to be updated.
Contact technical service department
Wait approximately 15 minutes for automatic
system reset
Check the supply system. If the problem persists,
contact the technical assistance department.
Repeat the x-ray. If the problem persists, contact
the technical assistance department.
Reduce the distance between the remote control
and x-ray head and then repeat the x-ray. Follow
the information given on how to properly use the
hand held's antenna. If the problem persists,
contact the technical service department.
X-ray generator or arm cord may be faulty.
Contact technical service department.
SOLUTION
the
problem
persists,
contact
EN