10. TECHNICAL DATA
EN
SPECIFICATIONS
-
Rated voltage: 230Vac/115Vac (according to the
model).
-
Max. mains voltage fluctuation: ±10%.
-
Maximum current: 6A for the 230Vac version;10A for
the 115Vac version at 60KV 7mA.
-
Frequency: 50/60Hz.
-
Maximum power absorbed: 1,4KVA.
-
Apparent
0,2Ω(115Vac).
-
Fuses: 6.3A T for the 230Vac version; 10A T for the
115Vac version.
-
Generator: constant potential type.
-
High nominal voltage: 60 / 63 / 65KV.
-
Maximum current: 6 / 7mA.
-
Power requirements at 0.1 sec: 420W (60KV 7mA),
378W (63KV 6mA), 390W (65KV 6mA).
-
Current reference time: 0.7 mAs (7mA – 0.1s) / 0.6
mAs (6mA – 0.1s).
-
Focal spot:
RX DC Plus: 0.4mm.
RX DC eXTend: 0.7mm.
-
Total filtration: 2.5mm Al @ 65KV.
-
Half-value layer (HVL): >2mm Al @ 65KV.
-
Leaked radiation: <0.25mGy / h at 1 metre from
focusing at 65KV 6mA,duty cycle 1:60.
-
Ability to be reproduced: 0.05.
-
Electrical classification: Class I - Type B, intermittent
service.
-
Set exposure time: from 0.010 to 1.000 seconds.
-
Accuracy of times indicated: ±10%.
-
mGy display precision: ±30%.
WEIGHTS
-
Weight of the unit with packaging: 38Kg (84lb) max.
-
Weight of the x-ray device: 25kg (55lb).
-
Weight of the hand-held control panel: 0,3kg (0.7 lb).
OPERATOR'S MANUAL
line
resistance:
0,5Ω(230Vac),
77