78
RADIOGENIC TUBE F.S.=0.4 mm
-
Radiogenic tube: TOSHIBA D-041.
-
Focal spot: 0.4 mm in compliance with IEC 336 /
1993.
-
Tolerance for position of the focal spot along the
reference axis: ± 2%.
-
Nominal high voltage and maximum allowable current:
(65KV, 7mA) ± 10%.
-
Anode construction material: Tungsten (W).
-
Anode inclination: 12.5°
-
Anode thermal load: 4.3 KJ (6 KHU).
-
Maximum continuous heat dissipation: 100 W.
-
Operating cycle: 1:60 (1 second exposure - 60
seconds pause time).
RADIOGENIC TUBE F.S.=0.7mm
-
Radiogenic tube: TOSHIBA D-0711.
-
Focal spot: 0.7mm in compliance with IEC 336 / 1993.
-
Tolerance for position of the focal spot along the
reference axis: ± 2%.
-
Nominal high voltage and maximum allowable current:
(65KV, 7mA) ± 10%.
-
Anode construction material: Tungsten (W).
-
Anode inclination: 16.0°.
-
Anode thermal load: 7.0 KJ (10 KHU).
-
Maximum continuous heat dissipation: 210 W.
-
Operating cycle: 1:60 (1 second exposure - 60
seconds pause time).
OPERATOR'S MANUAL
EN