3.3. CONTROL PAD

As illustrated in the figure below, the handheld has four function keys and a single x-ray emission key.
1
"Dentition area selection" key
2
"Body build selection" key
3
X-ray emission light
4
"Increase" key
5
"Decrease" key
6
"X-ray emission light" key
The main functions of the keys on the handheld vary according to how they are pressed:
KEY
BRIEFLY PRESSED (less than 3 sec.).
Changes over from ADULT to CHILD and
vice versa (takes place when key is
released).
It will change over the various types of teeth
to select the area to be examined.
Increases the exposure times in steps,
according to the set scale.
Decreases the exposure times in steps,
according to the set scale.
NO EFFECTS ARE OBTAINED IF THE
KEY
SECOND.
NOTE: "Dead man" function: the system that starts x-ray exposure with the dedicated key on the wireless
handheld allows x-rays to be emitted only when the user presses and holds down the exposure key. X-ray
emission will stop if the key is released ahead of time.
64
IS
PRESSED
LESS
OPERATOR'S MANUAL
PRESSED LONGER (more than 3 sec.).
Saves the selected setting (exposure time,
sensitivity, etc...). The memo icon(
when the data item can be saved.
It displays the values corresponding to the tooth
exposure times in mGy and in mGy*cm
pressed again.
Increases the scroll speed of the values in
increasing order.
Increases the scroll speed of the values in
decreasing order.
Starts x-ray exposure (the button has to be held
THAN
A
down while the x-rays are being emitted, "dead
man" function).
) lights up
if
2