3.5. FACTORY SETTINGS
The myray RX DC - RX DC/I x-ray unit is supplied with the following factory settings:
-
Operative mode: AUTO.
-
Sensitivty: level 19.
-
Handheld stand by: 5 minutes
-
Exposure times as per standard R'20: 0.010 - 0.011 - 0.012 - 0.014 - 0.016 - 0.018 - 0.020 - 0.022 - 0.025 - 0.028
- 0.032 - 0.036 - 0.040 - 0.045 - 0.050 - 0.056 - 0.063 - 0.071 - 0.080 - 0.090 - 0.100 - 0.110 - 0.125 - 0.140 -
0.160 - 0.180 - 0.200 - 0.220 - 0.250 - 0.280 - 0.320 - 0.360 - 0.400 - 0.500 - 0.560 - 0.630 - 0.710 - 0.800 - 0.900
- 1.000
NOTE: These times comply with current standards I.E.C. 60601-2-7 (1999) and the ISO 497 series R'20
recommendations and CANNOT BE MODIFIED.
3.6. BATTERIES AND CHARGE LEVEL INDICATION
The handheld runs on two widely available AA alkaline batteries to assure sufficient stand-alone operation.
The charge level of the batteries is given on the screen as follows:
Battery fully charged (a symbol does not appear in the area that shows the battery charge level).
Battery half-charged.
Battery charge level low or almost dead (causing the handheld to automatically shut off).
NOTE: The batteries should be removed from the handheld if it is not going to be used for an extended
period.
3.7. X-RAY GENERATOR INDICATOR LIGHT
In RX DC Plus versions, the x-ray generator comes with an indicator light (B) that signals
apparatus status.
Colors legend:
-
-
-
-
-
In the RX DC eXTend version, the indicator light is not available.
66
myray color (purple)
Flashing myray color (purple)
Blue
Yellow
Red
OPERATOR'S MANUAL
> x-ray unit on (regular condition)
> stand-by (low consumption)
> x-ray one – head released
> x-rays being emitted
> fault
EN