PROBLEM
Nothing appears in
the display, after the
(Power Button) is
pressed
Cuff does not inflate
Blood pressure
readings are too
high or too low
Blood pressure
readings are variable
CAUSE
Batteries are drained
Batteries are inserted incorrectly
Tube connector is not fully inserted into
device
Tube connector is broken
Cuff hose is compressed or restricted
Cuff is not fastened properly
Battery voltage is too low
Cuff is not fastened properly
Body and arm are not positioned properly
Movement or talking during measurement
Repeated measurement on same arm within
short time period
Possible factors: time of day, emotional
state, daily activities or exercise, smoking,
drinking alcohol or caffeine, eating, certain
medications (see pg. 6)
CORRECTION
Replace the batteries with new ones
Reinstall the batteries with positive and
negative terminals in correct position
(see pg. 13)
Firmly insert connector into air socket
on left side of device
Call product support
Make sure hose remains straight and
unkinked during measurement
Reposition the cuff correctly (see pg. 16)
Replace all batteries with new ones
Reposition the cuff correctly (see pg. 16)
Reposition your body and arm correctly
(see pg. 10)
Remain seated, and please refrain from
moving or talking during measurement
Wait 5 to 10 minutes between
measurements
Take measurement at same time each
day, and under the same conditions
21