7.7

Automatic cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be serviced
and checked at regular intervals.
7.7.1

Manual pre-cleaning with a brush

Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Push the saw blade coupling 14 towards the housing as far as it will go.
Phase I
Mobilize non-rigid components during cleaning, e.g. pull locking
sleeve 20 back from the support foot 16.
Thoroughly clean the product under running water.
Use cleaning template 15 to clean the slot of the saw blade
coupling 14. Insert cleaning template into the slot and carefully push
out the debris through the slot or the inspection holes at least 5 times
in both directions. Turn template 180° in order to do so.
Manually clean the contact sites of the rinsing adapter GA674890 and
the rotating sleeve 12 thoroughly.
Phase II
Mobilize non-rigid components during cleaning, e.g. pull locking
sleeve 20 back from the support foot 16.
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
10
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
-
-
RT (cold)
-
-
Water qual-
Chemistry/Note
ity
D–W
Until visually clean
D–W
Until visually clean