11. Accessories/Spare parts
Art. no.
Designation
GA674890
Rinsing adapter
GA675
Lid
GA676
NiMH battery long
GA678
Sterile funnel
GA679
Battery removal device
GB243R
ECCOS sieve basket with fixation for Acculan 4
GB244R
GB600
STERILIT Power Systems oil spray
GB436R
Support foot
GB440215
Cleaning template
GB495R
ECCOS fixation for large machine
GB487R
ECCOS fixation for lid
GB488R
ECCOS fixation for sterile funnel
GB489R
ECCOS fixation for battery removal device
GB686R
ECCOS support foot holder
TA014544
Instructions for use of reciprocating saw GA334
(A4 for ring binder)
TA014545
Instructions for use of reciprocating saw GA334
(flyer)
12. Technical data
12.1 Classification acc. to Directive 93/42/EEC
Art. no.
Designation
GA334
Reciprocating saw
12.2 Performance data, information about standards
Max. power
Max. oscillation frequency
Weight (operational)
Dimensions (L x W x H, ready for
operation)
Applied part
EMC
Conforming to standard
The product was subject to 500 preparation cycles as a test by the manu-
facturer and has passed.
Class
IIa
approx. 250 W
-1
15 000 min
1.54 Kg ±10 %
207 mm x 180 mm x 55 mm ±5 %
Type BF
IEC/DIN EN 60601-1-2
IEC/DIN EN 60601-1
12.3 Operating mode
Operating mode
12.4 Environmental conditions
Operation
Temperature
10 °C to 27 °C
Relative
30 % to 75 %
humidity
Atmospheric
700 hPa to 1 060 hPa
pressure
13. Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
►
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
14. Distributor in the US/Contact in Canada
for product information and complaints
Aesculap Inc.
3773 Corporate Parkway
Center Valley, PA, 18034,
USA
Operation with non-periodic load and
speed changes (type S9 pursuant to
IEC EN 60034-1)
■
30 second application, 30 second
pause
■
5 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Storage and transport
-10 °C to 50 °C
10 % to 90 %
500 hPa to 1 060 hPa
15