Chauvin Arnoux CA 6161 Guia De Inicio Rapido página 16

Ocultar thumbs Ver también para CA 6161:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 35
During continuity measurement, loop or line impedance measurement, differential test or dielectric test, the instrument
can generate high currents. Its internal temperature then rises.
When the instrument is too hot to operate properly, it indicates this by displaying a symbol in the status bar.
: The instrument temperature is high, but measurements are still possible.
!
: the temperature of the instrument is too high and measurements are no longer possible.
SAVING MEASUREMENTS
At the end of each measurement, you can save it by pressing
The memory is organized by sites, objects, Auto Scripts and measurements.
REVIEW OF RECORDINGS
Save each measurement in a site and an object.
The site represents the building where the object is located.
The object represents the machine to be tested.
16
.
To review measurements, start from
the home screen and press
.