* Specifications may be changed without notice in the event of improvement being
made.
1. Type of protection against electric shock: INTERNALLY POWERED
EQUIPMENT.
2. Degree or protection against electric shock: TYPE BF APPLIED PART.
3. Mode of operation: CONTINUOUS OPERATION.
4. Equipment not suitable for category AP&APG equipment use in presence.
STATEMENT
The system might not meet its performance specifications if stored or used outside
the temperature and humidity as mentioned below:
Operating conditions: +5℃~+40℃. 15%RH~93%RH
Storage conditions: -20℃~+55℃.
TROUBLESHOOTING
lf you have trouble in using the unit please check the following points first.
ERROR DISPLAY
Nothing is displayed When
you push the POWER
button or
Battery icon
flash
E1:can't normally Increase
pressure
E3 inflate pressure too high
E2E4: have shaking while
measurement
Battery icon on
0%RH~93%RH
POSSIBLE CAUSE
No battery installation
Battery worn out
The polarities of batteries
placed wrongly
Check your arm cuff if
any air leakage
Pressure value of more
than 299mmHg
Hand or body shaking
while measurement
Battery low power
70kPa~106kPa
HOW TO CORRECT
Insert batteries
Replace new batteries
Insert battery in the
correct polarities
Replace arm cuff with
new one
Re-measurement or
send back dealer for
re-calibrate pressure
keeping static and correct
gesture to measure again
Replace battery and
measure again