Table 4. Problems, possible causes, and possible solutions
Problem: Telemetry Status light stops flashing without a response after the
button is pressed.
Possible cause: Insufficient telemetry; the Patient Assistant did not
communicate with the cardiac monitor. Possible solution: Reposition the
Patient Assistant directly over the cardiac monitor, press the button again, and
wait for a response.
Possible cause: Electromagnetic interference. Possible solution: Move away
from the source of interference (an electrical or magnetic item).
Possible cause: Component failure. Possible solution: Contact your doctor or
clinic to replace the Patient Assistant.
Problem: Telemetry Status light turns a solid color, beeps, but no response
indicator appears in the display area.
Possible cause: The light in the display area is broken or there is a component
failure. Possible solution: Contact your doctor or clinic to replace the Patient
Assistant.
Possible cause: The light in the display area went on and off before you looked
at it. Possible solution: Look at the indicator display area sooner.