Table 4. Problems, possible causes, and possible solutions (continued)
Problem: The Patient Assistant continues to fail to communicate or respond.
Possible cause: The Patient Assistant is outside its temperature range: 49 °F
(9 °C) to 110 °F (43 °C). Possible solution: Move to warmer or cooler
surroundings.
Possible cause: Electromagnetic interference. Possible solution: Move away
from the source of interference (an electrical or magnetic item).
Possible cause: Batteries are low or depleted. Possible solution: Replace the
batteries.
Possible cause: The Patient Assistant needs to be replaced. Possible solution:
Contact your doctor or clinic to replace the Patient Assistant.
Battery replacement and disposal
When the Check Batteries indicator light is on (see Table 3 on page 27),
replace the batteries in your Patient Assistant as soon as possible (see
page 31). The Patient Assistant uses two N-size 1.5 V batteries. The
recommended battery type is alkaline manganese dioxide, type ANSI/
NEDA 910A and IEC LR1, which can be purchased at retail and camera
stores and from online merchants.
Follow local regulations for proper disposal of used batteries.
30 English