Protective Equipment
The pressure safety device (pressure relief with burst disc) protects the sys-
tem during a filter integrity test and during the emptying process with com-
pressed air against damage with overpressure of > 2.5 bar + safety tolerance
of 0.5 bar. The pressure safety device is mounted between the air filter and
the filter integrity tester or compressed air connection.
Fig. 7: Pressure Safety Device
Pos.
Name
1
Connection for e.g., filter integrity tester or air filter
2
Burst disc (installed inside)
1
1
2