Descargar Imprimir esta página

Sartorius Maxicaps MR Instrucciones De Manejo Originales página 27

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 137
1
Fix the lock for the swivel function in the filtration position (1).
t
1
Release the two Lifting Gear hooks from the filter unit attachment
t
points.
5.2
Mounting Valve Actuators
Three valve actuators are mounted on the filter unit:
— The inlet valve actuator on the top of the filter unit opens and closes the
inlet valve.
— The venting valve actuator on the top of the filter unit opens and closes
the venting valve.
— The outlet valve actuator on the bottom of the filter unit opens and clos-
es the outlet valve.
NOTICE
The internal valve on the filter unit may be damaged by mounting the valve
actuator multiple times!
Mount the valve actuator a maximum of three times on the same filter
t
element valve.
Requirement
The three valve actuators are in the [open] position.
Procedure
Remove the three valve actuators from the valve holder.
t
Filter Element with Venting Tubing Set
To mount the venting valve actuator on the top of the filter element:
t
Pull the valve actuator lock (1) upwards and place the valve actuator
t
on the top of the filter unit.
Turn the valve actuator clockwise until the lock engages.
t
Check that the valve actuator is mounted correctly. The valve actuator
t
must not turn counter-clockwise.
If required: Remount the valve actuator.
t
Close the valve actuator once and open it again.
t
The valve actuator connects to the valve membrane.
y
Process Preparation
Maxicaps
MR Operating Instructions
®
27

Publicidad