6.8
Replacing the batteries / rechargeable batteries
!
WARNING
Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in potentially explosive areas.
Do not use new batteries with used batteries, and do not mix batteries from
different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline
Varta Type 4106
Varta Type 4006
Alkaline batteries – T4 – (non rechargeable!)
Duracell Procell MN1500
NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC (type
HBT 0100) with the appropriate Dräger charger. Charge NiMH single cells
for ABT 0100 battery holder as directed by the manufacturer. Ambient
temperature during charging: 0 to +40 °C.
1.
Switching off the instrument: Press and hold the [OK] and [ + ] keys
simultaneously.
2.
Loosen the screw on the power pack and remove the power pack.
Battery holder (order no. 83 22 237): Replace alkaline batteries or NiMHy
rechargeable batteries. Ensure correct polarity.
!
WARNING
Explosion hazard!
The gas detector must only be operated with the ABT 0100 battery holder.
1)
Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
1)
(power one) or
1)
(industrial)
1)
1)
(1800 mAh) max. 40 °C ambient temperature.
Operation
en
41