Operation
NiMH power pack T4 (Type HBT 0000) / T4 HC (Type HBT 0100):
3.
replace complete power pack.
4.
Insert the power pack into the instrument and tighten the screw, the
instrument switches on automatically.
6.8.1
Charge instrument with NiMH power pack T4 (Type HBT 0000)/
T4 HC (Type HBT 0100)
!
WARNING
Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not charge underground or in explosion hazard areas! The chargers are not
designed in accordance with the regulations for fire damp and explosion protection.
Charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100) with
the appropriate Dräger charger. Ambient temperature during charging:
0 to +40 °C.
Even if the instrument is not used, we recommend that you store the instrument
in the charger!
Insert the switched off instrument into the charger module.
Display LED on the charger module:
To protect the battery charge only in the temperature range of 5 to 35 °C. Outside
this temperature range, the charging process is automatically interrupted and
automatically continued after the temperature range has been reached again.
The charging time is typically 4 hours. A new NiMH power pack reaches its full
capacity after three complete charge/discharge cycles. Never store the
instrument for extended periods without being connected to a power source
(maximum of 2 months) because the internal buffer battery will drain.
6.9
Carrying out the manual bump test
The automatic bump test with the Bump Test Station is described in the
Technical Manual.