MULTIPIEZO
05.4 Û INSTRUCTIONS FOR USE
After having connected all the accessories as illustrated in Paragraph 04.3 proceed as follows:
To fill the irrigation circuit use the PUMP
function by selecting PUMP/FLUSH on the
EN
touch keyboard: all the other selection
options presente on the display are disabled
and the PUMP/FLUSH writing flashes;
While the PUMP/FLUSH writing flashes,
press the foot pedal once and release it,
the PUMP/FLUSH stops flashing and the
irrigation circuit begins to be filled up;
As soon as the peristaltic pump starts, the
entire scale of values of the "irrigation"
section lights up and during the liquid
passage, the value of the irrigation shifts
from 6 to 0;
The cycle can be interrupted as soon you
see liquid outflowing from the multipiezo
handpiece, by pressing on PUMP/FLUSH
menu or, alternatively, by pressing the foot
pedal. The function PUMP is disabled and
the keyboard is enabled again, and displays
the last setting used;