08 Û CLEANING AND STERILIZATION
Clean the surface of the handpiece and
2
of its connector with a clean, soft cloth
with low fiber-release, dampened with
a detergent solution (pH 6-9) and, if
need be, disinfect with a non-aggressive
disinfectant solution with neutral pH (pH7),
following the instructions provided by the
manufacturer of the solution;
Delicately clean the surface of the
3
handpiece using the enzymatic detergent
solution with the aid of a brush with soft
nylon bristles, placing special care to the
following areas:
- threading of the handpiece
- titanium stem
- front terminal in its external and internal
parts
Accurately rinse with tap water to eliminate
4
any detergent residues and perform the last
rinse with distilled water;
Thoroughly dry all the parts, especially
5
the electrical contacts, blowing on them
with compressed air;