MULTIPIEZO
12.2 Û QUICK SOLUTION TO PROBLEMS
Problem
EN
The device does not turn
on after having brought
the switch into position
"I"
The device is on
but not working. The
display does not signal
any error
The device is on but
not working. One of
the following symbols
appears on the screen:
A slight whistling
sound coming from the
multipiezo handpiece
is heard
during operation.
44
Possible cause
The electrical power cable
terminal is badly inserted in
the rear plug of the device
The electrical power cable is
defective
The fuses are out of order
The foot pedal plug is
incorrectly inserted in the
device socket
The foot pedal does not work
See paragraph 12.1 for the
possible cause, according
to the symbol that has been
displayed
The insert is not correctly
tightened on the handpiece
The irrigation circuit has not
been completely filled
Solution
Check that the power supply
cable is firmly connected
Check that the power supply
socket works properly
Replace the electrical power
cable
Replace the fuses
(see paragraph 12.3)
Correctly insert the pedal plug
in the socket on the back of the
device
Contact an Authorized Mectron
Service Center
See paragraph 12.1 for the action
to undertake, according to the
symbol that has been displayed
Unscrew and correctly screw the
insert again with the Mectron
torque wrench
(see paragraph 05.4)
Fill the irrigation circuit via the
function FLUSH
(see paragraph 05.4)