MULTIPIEZO
Seal the wrench individually in a single-use
sterilization packet.
Sterilize the wrench in a steam autoclave.
EN
STERILIZATION METHOD - Torque wrench
The sterilization process in steam autoclave
must be performed setting the parameters
indicated below, which are validated by
Mectron and reach the sterilization safety
levels (SAL 10
-6
- 3 times Pre-vacuum
(minimum pressure 60 mBar).
- Sterilization temperature 132°C
(interval 0°C ÷ +3°C).
- Sterilization time 4 minutes.
- Minimum drying time 10 minutes.
All the sterilization phases must be performed
by the operator in compliance to the norms
UNI EN ISO 17665-1:2007, UNI EN ISO
556-1:2002 e ANSI/AAMI ST 46:2002.
08.6 Û CLEANING OF THE BOTTLE
PREPARATION
- Disconnect the bottle from the device body
(see Chapter 07);
- Unscrew the cap from the bottle.
MATERIALS NECESSARY
- Water
- Detergent solution (pH 6-9). If need be, non-
aggressive disinfectant solution with neutral
pH (pH7)
- Clean and soft cloth with low fiber-release
- Sterile water
34
):
8
WARNING: Checking infections - Parts
that can be sterilized - Diligently remove
all residues of organic dirt before the
sterilization.
CAUTION: Perform the sterilization
using a water steam autoclave only. Do not
use any other sterilization procedure (dry
heat, irradiation, ethylene oxide, gas, low
temperature plasma, etc.).
CAUTION: do not exceed the allowed
load of the steam sterilizer.
CAUTION: Do not sterilize the bottle
and its cap in autoclave, they could get
damaged.
CAUTION: If you intend to disinfect the
bottle and its cap we recommend that you
use water-based disinfectant solutions
with a neutral pH (pH7). Alcohol-based
disinfectant solutions and hydrogen
peroxide are contra-indicated, because
they can fade the color and/or damage the
plastic materials. This also holds true for
chemical products such as acetone and
alcohol. Always rinse with sterile water to
preserve the disinfection.