17
Figure 13: Single-use cable by Remington Medical Inc. (USA only)
Figure 14: Cable PK-155
WARNING!
WARNING!
WARNING!
Caution!
Note:
Note:
Cable for single use (USA only)
Cable for single use
The single-use cables PK-155 and Remington 301-CG
(USA only) with alligator clips (Figure 14) are connected to the
patient through the cable PK-67-S.
Connection
Danger to patient by damaged cables.
Damaged cables are limited in functionality and pose a dan-
ger to patients. Do not use damaged cables.
Danger from loss of function.
Damp cables have limited functionality and pose a danger to
patients. Do not use damp cables.
Danger from electrical currents.
Unused cable contacts can conduct electrical currents to
patients. Adhere unused cable contacts close to the patient.
Allergic reactions and inflammations.
Prevent the cable from coming into contact with the patient's
wounds or skin.
Ensure correct fitting of the insulators prior to using the
cables.
Do not connect the patient cable to the temporary pacing lead
of the patient before the connection has been established to
the Reocor D.
Direct connection
If Reocor D is used without the Redel adapter, temporary
catheters and heart wires can be connected to the patient
cables PK-82 and PK-83 directly at the connector ports A+, A-
and V+, V-.