Checking the resis-
tance to interference
Voltage drops, brief
interruptions and fluc-
tuations in the supply
voltage according to
IEC 61000-4-11
Magnetic field at the
supply frequencies
(50/60 Hz) according to
IEC 61000-4-8
Checking the resis-
Test level according to
tance to inter-
IEC 60601
ference
Conducted HF
10 V
interferences
10 kHz to 80 MHz
according to
outside of the ISM
IEC 61000-4-6
bands
10 V
10 kHz to 80 MHz
inside of the ISM
bands
Radiated HF inter-
10 V/m
ference according to
800 MHz to 2.5 GHz
IEC 61000-4-3
Test level
Compliance
according to
level
IEC 60601
Not applicable
3 A/m
30 A/m
Manufacturer guidelines and declaration – resistance to
electromagnetic interference for all external pacemaker
models (IEC 60601-1-2: Table 3)
The device is intended for use in an electromagnetic environ-
ment as described below. The user of the device should make
sure that it is used in such an environment.
Compliance
level
10 V
rms
a
10 V
rms
a)
10 V/m
Guidelines for the electromagnetic
environment
The magnetic field strength should
correspond to the typical value in
business and hospital environ-
ments.
Guidelines for the electromagnetic
environment
Portable and mobile radio devices
are not used closer to any part of
the device, including cables, than
the recommended safe distance.
Recommended safe distance:
rms
d = 0.35
rms
d = 1.2
d = 1.2
P
for 80 MHz to 800 MHz
d = 2.3
P
for 800 MHz to 2.5 GHz
P
P