EN – ENGLISH
Table I Guidance and Manufacturer's Declaration – Emissions
All ME Equipment and ME Systems
The 823190R HAKIM Programmer is intended for use in the electromagnetic environment specified below. The customer or user of the 823190R HAKIM Programmer should
ensure that it is used in such an environment. Deviations from emission and immunity environment for this device may affect its life expectancy.
Emission Test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonic Emission
IEC 61000-3-2
Flicker
IEC 61000-3-3
Table II Guidance and Manufacturer's Declaration – Immunity
All ME Equipment and ME Systems
The 823190R HAKIM Programmer is intended for use in the electromagnetic environment specified below. The customer or user of the 823190R HAKIM Programmer should
ensure that it is used in such an environment. Deviations from emission and immunity environment for this device may affect its life expectancy.
Immunity Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips/Dropout
IEC 61000-4-11
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
Compliance
Electromagnetic Environment – Guidance
Group I
The 823190R HAKIM Programmer uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Class A
The 823190R HAKIM Programmer is suitable for use in all establishments, other than domestic, and those
Class A
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Complies
IEC 60601
Compliance
Test Level
Level
±8kV Contact
±8kV Contact
±2kV, ±4kV, ±8kV, ±15kV air
±2kV, ±4kV, ±8kV, ±15kV air
±2kV Mains
±2kV Mains
±1kV I/Os
±1kV I/Os
±1kV Differential
±1kV Differential
0 % UT; 0,5 cycle UT 0°, 45°,
0 % UT; 0,5 cycle UT 0°, 45°,
90°, 135°, 180°, 225°, 270° and
90°, 135°, 180°, 225°, 270° and
315°
315°
0% UT; 1 cycle at 0°×0 % UT;
0% UT; 1 cycle at 0°×0 % UT;
25/30 cycles 0°
25/30 cycles 0°
0% UT; for 5 seconds
0% UT; for 5 seconds
30 A/m
30 A/m
Electromagnetic
Environment – Guidance
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the
r/h should be at least 30%.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the 823190R HAKIM Programmer requires
continued operation during power mains interruptions, it is recommended
that the 823190R HAKIM Programmer be powered from an uninterruptible
power supply or battery.
Power frequency magnetic fields should be that of a typical commercial or
hospital environment.
6