Page E-20/E-24
Technical data
Manufacturer:
Model:
Classification:
Mode:
Supply voltage:
Mains frequency:
Power consumption:
Fuses:
Compressed air supply:
Water supply:
Weight:
Dimensions:
Operating conditions:
Storage and transport conditions: -10°C - +40°C,
EMS SA, CH-1260 Nyon, Switzerland
AirFlow S2
EN 60601-1:
- Class I
- Type B
- IP 20
93/42 EU:
- Class IIa
Continuous operation
100 V AC / 110 V AC
220 V AC
240 V AC
50 / 60 Hz
max. 80 VA
Scaler = 14 VA
27 - 30 kHz
∅ 5 x 20 mm, 250 V AC, time lag
for 100 V AC / 110 V AC power supply:
for 220 V AC / 240 V AC power supply: T 0,5 A L
4.5 - 7.0 bars (450 - 700 kPa) > 15 Nl/min
Cold water 1.0 - 5.0 bars (100 - 500 kPa)
Approximately 4 kg
Height:
120 mm
Width:
250 mm
Length:
250 mm
+10°C - +40°C,
30% - 75% relative humidity
10% - 95% relative humidity,
500 hPa - 1060 hPa air pressure