• It is the responsibility of the Healthcare Professional to determine the appropriate uses of this
product and to understand:
○ The health of each patient
○ The dental procedures being undertaken
○ Applicable industry and governmental agency recommendations for infection control in dental
healthcare settings
○ Requirements and regulations for safe practice of dentistry
○ These Instructions for Use in their entirety.
• Failure to follow recommendations for environmental operating conditions (see section 6.3) could
result in injury to patients or users.
• Inspect equipment before each use for worn, loose or damaged parts.
• There are no user serviceable parts except the O-ring attached to the coupling end of the
attachment. Opening any of the components may result in unsafe operation and will void
the warranty.
• According to IEC60601-1, this device must not be used in the presence of a flammable anesthetic
gas mixed with air, oxygen or nitrous oxide. (Note: nitrous oxide by itself is not a flammable
anesthetic gas).
• User should not touch the patient and accessible charging base contacts or USB contact
simultaneously.
• Wear suitable protective eyewear, mask, clothing and gloves. Protective eyewear is recommended
for patients.
• Devices marked "single-use" on the labeling are intended for single-use only. Discard after use. Do
not reuse in other patients in order to prevent cross contamination.
• As a precautionary measure the handpiece and attachment may be protected from gross debris,
but not all contamination, by applying a protective barrier sleeve. Reprocess reusable components
after each use according to instructions.
• Do not spray disinfectant or other fluid directly onto the handpiece, attachment, battery, charging
base, power supply or cord. The user should spray solution onto a cloth or use a wipe to disinfect
items per instructions in Section 4.
• Prevent liquids from entering the handpiece, battery pack, and charging base.
• Ensure the battery contacts are fully dry before charging batteries or attaching batteries into the
handpiece to prevent corrosion. Similarly, ensure that the contacts on the attachment are fully dry
before attaching to the handpiece body.
• Do not place the system on or next to a radiator or other heat source. Excessive heat may damage
the system's electronics.
Interactions
None Known.
2.3 Adverse Reactions
None Known.
2.4 Storage Conditions
Inadequate storage conditions may shorten the shelf life and may lead to malfunction of the
product.
• Store at temperatures between -5°C/35°C (23°F/95°F).
• Use the product at room temperature.
• Protect from moisture.
• Store at relative humidity range <75% (non-condensing).
• Do not use after expiration date.
3. STEP-BY-STEP INSTRUCTIONS
SmartLite Pro EndoActivator™ – Operation at a glance
ON/OFF button
First Button Press – Turns the unit On and starts the activation of the activator tip
Second Button Press – Reduces the speed of the activator tip
Third Button Press – Turns the unit Off.
10