PACKAGES
Fig. 4
REF IP1
SINGLE LUMEN INTRODUCER KIT
Packaged in a sterile double tray. When REF IP1 is
supplied as kit with REF CC1P1 this kit is protected
by an outer plastic bag.
This outer plastic bag is non-sterile. It serves
to protect the inner packages from moisture
during shipping and storage.
Please remove and discard the plastic bag prior
to use.
a
Introducer (Ø 1.9 mm)
a1 Connector for combined oxygen and
temperature probe (Ø <1.4 mm)
a2 Compression seal
a3 End of lumen for combined oxygen and
temperature probe
b
Guide wire
c
Compression cap
d
Bolt
e
Drill bit (Ø 3.8 mm)
f
Adjustable drill stop with set screw
g
Hex wrench for adjustment of the drill stop
h
Stylet
Fig. 5
REF CC1P1
COMBINED OXYGEN AND TEMPERATURE
PROBE
For Oxygen Pressure (p
Cerebral Tissue.
Packaged in a sterile double tray protected by an
outer plastic bag.
This outer plastic bag is non-sterile. It serves
to protect the inner packages from moisture
during shipping and storage.
Please remove and discard the plastic bag prior
to use.
The probe (length 462 mm) [c, d, e] is delivered
(as shown on the right [a, b]) in a closed protection
tube [b], which provides mechanical protection and
protects the probe from drying out. The calibration
data for the probe are electronically stored on a
smart card [g]. Probe and smart card have the
same number. The extension tube [d] serves to
connect the probe to the introducer system. The
probe is connected at the rear end [c] of its plug to
the probe cable REF PMOCAB.
At [e] the probe has its p
proximately 18 mm
(Ø 0.65 mm). At [f] the probe
2
has its temperature sensitive area of approximately
23 mm
(Ø 1.3 mm). The entire protection tube
2
must be removed at [a] before use.
O
) Monitoring in
bt
2
O
sensitive area of ap-
bt
2
5