Applying Blood Sample
Step 7
Obtain a blood sample using the lancing device.
Place the lancing device against the pad of the
finger. The best puncture sites are on the middle
or ring fingers.
Press the release button. Remove the lancing device from the
finger. Wait a few seconds for a blood drop to form.
A minimum volume of 0.5 μL is needed to fill the confirmation
window. (Approximate size of 0.5 μL:
Step 8
After the
sample to the narrow end of the test strip until the meter beeps.
If the confirmation window is not filled in time because of
abnormal viscosity (thickness and stickiness) or insufficient
volume, the Er4 message may appear.
It is recommended to place the test strip vertically into the blood
sample site as shown below.
NOTE
• The meter may switch off if the blood sample is not
applied within two minutes of the
appearing on the screen.
• If the meter turns off, remove the test strip, reinsert it, and
start from step 2.
36 www.arkrayusa.com
symbol appears on the screen, apply the blood
)
Good
Sample
symbol