Calculation of separation distance for non-life-
Maximum
electrical
output
of trans-
mission
device (W)
0.01
0.1
1
10
100
Important information on electromagnetic compatibility
(EMC)
A high number of electronic devices such as PCs, mobile tele-
phones, etc., may cause electromagnetic disturbances in the
use of medical electronic devices. Electromagnetic disturbanc-
es may lead to failure in medical devices and possibly result in
dangerous situations. Medical products must also not lead to
faults in other devices.
The standard IEC60601-1-2 was introduced in order to meet
EMC (electromagnetic compatibility) requirements and pre-
vent dangerous situations with such devices. This standard
specifies the immunity of devices to electromagnetic interfer-
ence and the maximum values for electromagnetic emissions
for medical devices.
supporting systems
(3Vrms / 3 V/m conformity)
Distance depending on frequency of transmis-
150 kHz to 80
Mhz d=1.2
0.12
0.37
1.17
3.69
11.67
Electrical interference immunity
sion device (m)
80 Mhz to
800 Mhz
d=1.2
0.12
0.37
1.17
3.69
11.67
800 Mhz to
2.7 Ghz d=2.3
0.23
0.77
2.33
7.38
23.33