Steam Sterilization - Braun Aesculap AdTec mini Instrucciones De Manejo/Descripción Técnica

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 14
Phase III
Fully immerse the product in the disinfectant solution.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.
Rinse lumens with an appropriate disposable syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.
Mechanical cleaning/disinfecting
Note
The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
T
[°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
III
Intermediate
>10/50
rinse
IV
Thermal disin-
90/194
fecting
V
Drying
-
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Check visible surfaces for residues after mechanical cleaning/disinfecting.
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual pre-cleaning with a brush
Phase
Step
T
[°C/°F]
I
Disinfec-
RT (cold)
tant
cleaning
II
Rinsing
RT (cold)
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
t
Water
Chemical/Note
quality
[min]
3
D–W
-
10
FD-W
Concentrate, alkaline:
– pH = 13
– <5 % anionic surfactant
0.5 % working solution
– pH = 11*
1
FD-W
-
5
FD-W
-
-
-
According to the program for cleaning
and disinfection device
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
D–W
-
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
T
[°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
III
Intermediate
>10/50
rinse
IV
Thermal disin-
90/194
fecting
V
Drying
-
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Check visible surfaces for residues after mechanical cleaning/disinfecting.
Inspection, maintenance and checks
Damage (metal seizure/friction corrosion) to the product caused by insufficient
lubrication!
Prior to function checks, lubricate moving parts (e.g. joints, pusher compo-
nents and threaded rods) with maintenance oil suitable for the respective ster-
CAUTION
ilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray
JG600 or STERILIT® I drip lubricator JG598).
Allow the product to cool down to room temperature.
After each complete cleaning, disinfecting and drying cycle, check that the instrument is dry, clean, operational,
and free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured compo-
nents).
Dry the product if it is wet or damp.
Repeat cleaning and disinfection of products that still show impurities or contamination.
Check that the product functions correctly.
Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service.
Assemble dismountable products, see Assembling.
Check for compatibility with associated products.
Packaging
Appropriately protect products with fine working tips.
Store products with ratchet locks fully opened or locked no further than in the first notch.
Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected.
Pack trays appropriately for the intended sterilization process (e.g. in sterile Aesculap containers).
Ensure that the packaging provides sufficient protection against recontamination of the product during storage.

Steam sterilization

Note
The product can be sterilized either in disassembled or in assembled condition.
Note
To avoid breakage due to stress crack corrosion, sterilize the instruments with the lock fully open or locked no further
than on the first ratchet tooth.
Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets).
Validated sterilization process
– Steam sterilization using fractionated vacuum process
– Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding time 5 min
When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capac-
ity of the steam sterilizer specified by the manufacturer is not exceeded.
Sterilization for the US market
Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization.
Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave.
-6
To achieve a sterility assurance level of 10
, Aesculap recommends the following parameters:
Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*
Sterilization method
Temp.
Prevacuum
270 °F/275 °F
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an
Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles
may also be suitable, however individuals or hospitals not using the recommended method are advised to validate
any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility
after processing, such as a wrap, pouch, etc.
Storage
Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
t
Water
Chemical
quality
[min]
3
D–W
-
10
FD-W
Concentrate, alkaline:
– pH = 13
– <5 % anionic surfactant
0.5 % working solution
– pH = 11*
1
FD-W
-
5
FD-W
-
-
-
According to the program for cleaning
and disinfection device
Time
Minimum drying time
4 min
20 min