High Frequency Patient Cable
Attach one end of high frequency patient cable to output connector on Generator front panel. Line up the dot on the cable connector with
the corresponding dot on the front panel connector. Push the cable connector into place. Assure that the connection snaps into place.
Patient Cable and Transducer
Connect other end of patient cable to the connector at rear of Transducer, using the same procedure as above. Assure that the connection
snaps into place.
CAUTION: Do not hold the probe tightly while the power is on. Clamping the probe restricts the longitudinal motion, causing the probe
to heat at that point. Prolonged restriction may cause the probe to break. Restriction may also cause a mechanical
limit condition, causing ultrasonic energy to be inhibited. Undue bending of the probe may do the same. If both
these conditions are relieved, the mechanical limit condition will be cleared.
The AutoSonix™ System is now ready for system test.