6
OPeRATing gUiDelines
Connect the (sterile) distal end to the patient.,
We recommend connecting the patient-end of the tubing hose to a Veress needle in the first
phase. This first phase corresponds to low-flow insufflation (max. 2L/min) until the pressure set-
point is reached. Once the pressure setpoint has been reached, you can move on to the second
phase consisting of high-flow mode (up to 45L/min) with the patient-end tubing hose connected
to the trocar.
The use of a Veress needle is not recommended for high-flow mode.
This operation must be carried out by a surgeon or under his supervision.
This insufflator is equipped with automatic flow rate regulation to attain and maintain a pressure
setpoint, and therefore the pneumoperitoneum
Use the " + " and " - " buttons to select the pressure desired in the abdominal cavity.
The pressure generally used is 12 mm Hg.
As a precautionary measure, you will be asked to confirm the pressure if you wish to exceed
the 15 mm Hg threshold.
The pressure can be adjusted to a maximum of 20 mm Hg in Standard mode, and to 25 mm
Hg in bariatric mode.
Press the "RUN" button on the interface menu to start insufflation. A chronometer is displayed
on the screen when insufflation is in progress.
is in progress. To access it, insufflation must be halted by clicking the "STOP" button.
Insufflation begins in low-flow mode (2 l/min), so that the pneumoperitoneum is created in the
safest of conditions. When the low flow rate is activated, the "low flow rate" button on the in-
terface is displayed in blue.
The high flow rate mode is activated manually by pressing the corresponding button on the
interface menu or automatically if automatic change-over is activated (in the menu parameters).
Automatic change-over is used to automatically change from low-flow mode to high-flow mode
once the pressure setpoint has been reached. When the high flow rate is activated, the "high
flow rate" button is displayed in blue.
6.3 ADJUsTing The PRessURe seTPOinT
6.4 ACTiVATing / sWiTChing-OFF insUFFlATiOn
The MENU is not accessible when insufflation
- 19 -