Aesculap
Lektrafuse HF Generator GN200
Validated cleaning and disinfection procedure
Validated procedure
Wipe disinfection for electrical
devices without sterilization
5.6
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
5.7
Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
►
Set aside the product if it is damaged.
►
Perform a safety inspection annually, see Maintenance.
5.8
Storage and transport
Damage to the product caused by switch-on too
early after storage/transport at temperature below
+10 °C (50 °F)!
CAUTION
►
Allow the HF device to acclimate at room tem-
perature for about 1 hour.
►
Transport the product only in its original box.
►
For storage and transport conditions, see Ambient conditions.
12
®
Special features
–
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
6.
Maintenance
No servicing is required for the Lektrafuse HF generator GN200.
Note
Maintenance must only be carried out by authorized personnel. If neces-
sary, the circuit diagrams and the service manual, which contains all the
necessary documents, are provided for this.
6.1
Safety inspection
A safety inspection must be carried out annually.
The person in charge of the inspection will document the inspection find-
ings and measurement values according to the printed inspection proto-
col.
►
Only have the product and its accessories inspected by persons who
possess the requisite training, knowledge and experience and who are
authorized to do so.
►
If there are significant deviations from the inspection protocol values,
or if the prescribed maximum values are exceeded: Send in the device.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Reference
Chapter Wipe disinfection for electrical devices
without sterilization